CONTACT US

RADIOMETER TRANSCUTANEOUS CO2 Monitor, NON-INVASIVE

TCM5 – A non-invasive way to monitor oxygenation (tcpO2) and
ventilation (tcpCO2) status in neonatal, paediatric, and adult patients.

Focus on patient care

With its advanced features, the TCM5 FLEX helps support your workflow so
you can focus on patient care.

  • System defined limits of site time vs. temperature
  • Easy workflow
  • Built-in tutorials and clear how-to messages
  • Colour touchscreen for a clear view of measured parameters
  • Connectivity to major patient monitoring system

Make informed decisions quickly and with confidence

Continuous non-invasive monitoring helps keep you in control.
The TCM5 FLEX monitor gives you a real-time view of changes.

  • Colour touchscreen displays parameters continually
  • Real-time trending measurements enable immediate clinical
    intervention to critical changes in patient status
  • Clearly visible alarm bar on top of the monitor

Reliable sensors for accurate results

  • Surgical steel protected sensors help ensure accurate results
  • A variety of fixation options support patient comfort
  • Automatic calibration of the sensor reduces preparation time and improves workflow
  • New for 2022: Improved sensor head design that with better protection of the cable and greater pull strength

Operating conditions

  • Ambient temperature: 5-40 °C (41 – 107 °F)
  • Relative humidity: up to 90 %, non-condensing
  • Atmospheric pressure: 525 – 800 mmHg (700 – 1060 hPa)
  • Built-in barometer:
  • Range: 375-825 mmHg or 50-110 kPa
  • Accuracy: ±5 mmHg or 0.67 kPa
  • Power: 100-240 V, 50-60 Hz, 55-85 VA (max.)
  • Rechargeable Lithium Ion battery, Type RRC2040
  • Typical operating time: 4 hours per charge at 25 °C

DOWNLOAD BROCHURE

DOWNLOAD

DOWNLOAD SPECIFICATION

DOWNLOAD

DOWNLOAD ADDITIONALS

DOWNLOAD
DOWNLOAD
  • Sensor performance (in vitro)

    tc Sensor 92

    • Measurement principle:
      Stow-Severinghaus-type tcpCO2 sensor combined with Masimo SET® SpO2 pulse oximetry. Conditions: sensor temperature of 43.0 °C, calibration interval: 12 h
    • tcpCO2:
      Accuracy: (5-200 mmHg): ± 5 mmHg
      Response time: (0-90%): ≤ 70 sec
      Drift: ≤ 0.5%/h
      Linearity: at 1 and 10% CO2: better than 1 mmHg or 0.13 kPa
      at 33% CO2: better than 5 mmHg or 0.67 kPa
    • SpO2:
      Accuracy: (70-100%): ± 3 digits
    • Pulse rate:
      Accuracy: (25-240 bpm) ± 3 bpm
      SpO2 and pulse rate: signal averaging 2, 4, 8, 10, 12, 14 and 16 sec
      Data update rate: 1 per second

     

    tc Sensor 84

    • Measurement principle:
      Stow-Severinghaus-type tcpCO2 sensor combined with Clark-type tcpO2 sensor
      Conditions: sensor temperature of 43.0 °C, calibration interval: 12 h
    • tcpCO2:
      Accuracy: (5-200 mmHg): ± 5mmHg
      Response time: (0-90%): ≤ 70 sec
      Drift: ≤ 1%/h
      Linearity: at 1 and 10% CO2: better than 1 mmHg or 0.13 kPa,
      at 33% CO2: better than 5 mmHg or 0.67 kPa
    • tcpO2:
      Accuracy: (0-20.9 % O2): ± 5 mmHg (20.9-100 % O2): ± 10 % Response time: (0-90%): ≤ 25 sec Drift: ≤ 1%/h Linearity: at 0% O2: better than 1 mmHg or 0.13 kPa at 20.9% O2: better than 3 mmHg or 0.4 kPa at 50% O2: better than 5 mmHg or 0.67 kPa at 90% O2: better than 25 mmHg or 3.33 kPa

     

    tc Sensor 54

    • Measurement principle:
      Stow-Severinghaus-type tcpCO2 sensor
      Conditions: sensor temperature of 43.0 °C, calibration interval: 12 h
      tcpCO2: Accuracy: (5-200 mmHg): ± 5 mmHg
      Response time: (0-90%): ≤ 70 sec
      Drift: ≤ 0,5%/h
      Linearity: at 1 and 10% CO2: better than 1 mmHg or 0.13 kPa
      at 33% CO2: better than 3 mmHg or 0.40 kPa


    Info Contact

    HEAD OFFICE
    Tel +27 (11) 966 0600
    Email info@medhold.co.za
    Address MSI Business Park, 68 Rigger Road,
    Spartan, Kempton Park,
    Gauteng, 1619

    Cookies preferences

    Others

    Other uncategorized cookies are those that are being analyzed and have not been classified into a category as yet.

    Necessary

    Necessary
    Necessary cookies are absolutely essential for the website to function properly. These cookies ensure basic functionalities and security features of the website, anonymously.

    Advertisement

    Advertisement cookies are used to provide visitors with relevant ads and marketing campaigns. These cookies track visitors across websites and collect information to provide customized ads.

    Analytics

    Analytical cookies are used to understand how visitors interact with the website. These cookies help provide information on metrics the number of visitors, bounce rate, traffic source, etc.

    Functional

    Functional cookies help to perform certain functionalities like sharing the content of the website on social media platforms, collect feedbacks, and other third-party features.

    Performance

    Performance cookies are used to understand and analyze the key performance indexes of the website which helps in delivering a better user experience for the visitors.